The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 24
The Interval time level is based on how long a unit can run without AC power, 99 seconds is the maximum
transmission time interval.
To check if the DRM is in Power Saving Mode press on the ‘GET BUTTON’ in the “Setup Device” window in
the RMV software and the background window will be display ‘8888’ to signify that the DRM is in the Power
Saving Mode.
4.2.10 Dose Rate Alarm Threshold Settings
Dose threshold settings are configured via the RMV software (Version 4.01 and DRM firmware ). The Def.
T
hreshold (thr.0) is the current default setting in the DRM, i.e. Threshold (1) one is the current default dose rate
threshold (see figure below), thr.0. To change the threshold setting using the ACK button on the display, toggle
to the desired Threshold number, i.e. thr.3, let go of the ACK button and the display will show the stored
threshold setting and will be defaulted to this rate threshold alarm setting.
Example of the Setup Device Window for with the Dose Rate Threshold Settings
4.2.11 Dose Function
Dose reading can be read as a part of the data record sent by the DRM.
To enable this function:
1) Connect the DRM to the PC and run RMV software
2) Move dipswitch 1&2 to the ON position, and set DOSE enable in the RMV software
If the DOSE reading is disabled, in place of a real dose reading, the record will be shown as 888888.8. The Dose
reading will be reset to zero in two cases: Disable\Enable Dose function, turn the DRM On. Each time the
instrument is turned ON; the Dose reading is reset to zero.