10.5 Calibration
Select Instrument/Calibrate… from the pull down menu. The calibration menu will be displayed.
(1) Instrument reading – Current dose rate reading
(
2) Factor Num. - Factor number for the calibration. The listing in this screen depends on the amount of
calibration points available from the instrument. If the “Auto” radio button is selected the software will recognize
the filed value and allow for calibration of the pertinent point. If the calibrator wants to insert a factor into a
specific range, he or she simply selects the field.
(3) Factor Num. - Current active factor number. This window will display the current active factor number
and is especially useful in the Auto mode for reference purposes.
Note: The calibration factors are selected according to the detector's radiation field
location. Only the displayed factor can be changed.
(4) Old Factor: - Previous calibration factor as saved (stored) in the instrument memory.
(5) Enter Radiation Field: - Factor can be changed in two modes:
a
. Insert the measured field. The computer will calculate the new factor dependant on the radiation field
and the old factor. The new factor value will be displayed on the New Factor box.
b. Using the ± boxes.