Seitenansicht 33
15-00043
Revision 3
September 2005
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 34
APPENDIX 1
Dosimeter Data Protocol
This document describes the data exchange protocol used between a DMC series dosimeter in 4800 BAUD
and a PAM-TRX sent to a Base or Repeater.
Dosimeter Settings
The correct dosimeter settings for the dosimeter are as follows:
Status word 5 (accessible from Message 11)
Bit 0 set to 1 tele-transmission allowed
Bit 2 set to 1 transmit on request mode
Bit 3 Set to 1 transmit in 4800 BAUD (DMC-2000 only)
This means that the dosimeter transmits data immediately following a strobe signal and the data is
transmitted using the dosimeter internal clock in ASCII format, 8 bits with bit 8 set to 0, and 1 stop bit.
Figure A1 below depicts a timing diagram of the data stream.
Signal Out bo b1 b2 b3 b4 b5 b6 b7
LSB MSB
START STOP
Figure A1-1: Data Stream Timing Diagram
The ASCII format of the data is as shown in Table 1 below. The dosimeter data consists of 13 characters,
each consisting of two bytes representing an ASCII character or a set of 4 status bits. The data starts with a
Line Feed (LF) and ends with a checksum CKS. A Carriage Return (CR) is transmitted after the checksum.
The data sent is:
Dosimeter serial number
Cumulated Dose
Current value of the Dose Rate
Dosimeter Status
The check sum is calculated such that the modulo 256 sum of all 26 bytes is equal to 0.