TelePole WR/ Operating & Maintenance Manual Page 20
The CPU unit counts the input pulses, calculates dead time, averages the results, and displays
the reading.
To check the CPU unit "calibration" perform the following procedure:
5.3.1. Disconnect the detector telephone connector from the meter.
5.3.2. Adjust the pulser output to obtain a 5 Volts amplitude and 10
µ
µµ
µ
sec width pulse.
5.3.3. Connect the pulser output as follows: (+) to pin 1 and (-) to pin 2.
Option: A more convenient and easier way to perform this step is by using the TelePole
Pulser Adapter. Proceed as follows:
- Connect the adapter telephone cable to the detector connector (Det.) on the TelePole.
- Connect the pulser signal output to pin 4 on the adapter board.
- Connect the pulser ground output to pin 3 on the adapter board.
5.3.4. First, turn the meter on, and then turn the pulser on. Ensure to obtain the base unit
reading (±10%) according to the following table:
Tab1e 1 - Low Range Geiger
Input Frequency <Hz>
Display <mR/h>
17 1.00
170 10.0
2000 141
4000 365
5000 534
6000 730
6300
804*
* Switch to high range geiger
Table 2 - High Range Geiger
Input Frequency <Hz>
Display <R/h>
300 1.00
3200 10.7
10.000 39.2
30.000 183
40.000 323
50.000 525
60.000 839
75.000 O.F. (overflow)
These results are valid only when all factors equal to 1.