MGP WRM2 Spezifikationen Seite 28

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 58
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 27
TelePole WR/ Operating & Maintenance Manual Page 28
Internal / External Geiger Selection
U2 - 74HC132 and port J2/4 are used to select between the internal and external geiger.
J2/4 enables to select between operation and display of internal geiger or external geiger.
J2/4 = 0 - internal geiger
J2/4 = 1 - external detector.
3V to 5V Converter & Low Battery Identification
MRX756 (U1) and RCL components convert the battery voltage to 5V, (which is) the
meters main power supply.
Input voltage range: 1.7V to 3V DC.
Output voltage range: 5V ±0.1V.
Low battery identification is performed by U1. When the battery voltage reduces below
2.0V on pin LB1, LB0 (pin 4) on U1 resets to 0.
3V to 12V Converter
MAX771 (U4) and RCL components convert the battery voltage to 12V that is aimed for
the speaker circuit.
7.2. WRM Communication
Q2, Q3 and resistors adapt between the WRM transmitter and meter voltage levels.
E1 (1-3) should be shorted.
7.3. Wide Range Detector
DRW # 12852-40-00
Detector Block Diagram
Pulse
Shaper
H.V.
GM 1201
Low
Range
H.V.
Power supply
500V
±
±±
±
20V
Select
Geiger
Driver
Pulse
Driver
Pulse
Shaper
H.V.
GM 1301
High
Range
J1
Seitenansicht 27