2. Specifications
Detector GM tube ZP-1301 (or equivalent) - high range
GM tube ZP-1201 (or equivalent) - low range
Measuring range 0.05 mR/h to 1000 R/h (0.5 µSv/h to 10 Sv/h)
Automatic switching between the two GM tubes at
approximately 800 mR/h and 600 mR/h, optional
firmware to switch 2500 mR/h and 2000 mR/.h.
Accuracy ±10 % of reading, within the measuring range
Energy response (Cs-137) ±20 % at 70 keV to 1.1 MeV
Sensitivity (Cs-137) Approximately 17 cps/mR/h (low range)
Approximately 0.3 cps/mR/h (high range)
Data logging 380 data records (1550 with extended memory)
Display DigiLog (3 digits and 2 decades of analog bar graph)
Power source Two 1.5 Volt C-type alkaline cells,
70 hours of continuous operation
(30% less if back light is kept on continuously)
Temperature range Operation: -10°C to +50°C (15°F to 122°F)
Storage: -20°C to + 60°C (-5°F to 140°F)
Humidity range 10% to 95% RH (non condensing)
Casing Material Aluminum, black anodizing
Dimensions Meter Width: 93 mm (3.7”)
Length: 148 mm (5.8”)
Height: 56 mm (2.2”)
Probe head Length: 226 mm (8.9”)
Diameter: 33 mm (1.3”)
Pole length Extended: Up to 337 cm (11 ft)
Pole length Collapsed: Down to 107 cm (3.5 ft)
Weight Meter 680 gr. (1.5 lbs.)
Probe 175 gr. (0.39 lbs.)
Pole 1200 gr. (2.7 lbs.)
Total 2055 gr. (7.29 lbs.)