TelePole WR/ Operating & Maintenance Manual Page 29
High Voltage Power Supply
The H.V. board includes an ORAM 5525 transformer, a voltage multiplier (D1, D2, D3,
C8, C11), an RV4192 switching regulator.
R3 and R4 are voltage feedback for the switching regulator.
H.V. range: 500V to 550V DC. The high voltage is aimed for both geigers.
Pulse Shaper
Each geiger includes a pulse shaper circuit.
Q3 and RC for the low range geiger 1201.
Q5 and RC for the high range geiger 1301.
7.3.1. Select Geigers
The detector includes two geigers: Low range - ZP-1201 (or equivalent), and high range -
ZP-1301 (or equivalent).
In a field of 0.01 mR/h to 800 mR/h both geigers are connected and the field is measured by
the low range geiger. In case the radiation field is higher than 800 mR/h, the low range
geiger is disconnected, and the radiation field is measured by the high range geiger.
The switching between the low and high range geigers is done at 800 mR/h when the
radiation field increases (low range geiger is disconnected), and at 600 mR/h when the
radiation field decreases (low range geiger is connected).
Geigers selection is performed via J1/2 control line.
Pulse Drive
Q2, Q4 and R23 drive the detector pulses to the meter.
High Range
Geiger Type
Low Range
Overflow
alarm
Field
800 mR/h
1000 R/h
ZP 1301
ZP 1201
600 mR/h
0.01 mR/h
Seitenansicht 28